Service Bulletins Service Bulletins are published to address common questions and concerns that may arise during installation and use.
Advisory Notices are issued when potential problems are discovered that may affect already installed equipment, but do not require recall.
The ConRad / ZooMax generators do not have any potentiometers to adjust kV output. If necessary, the kV may be adjusted by software in Service Mode.
Related to: ConRad HF, ZooMax HF
Some Milestone HF generators may turn off intermittently during or between exposures regardless of operating mode or conditions. The generator can then be turned on again using the ON switch on the console. The problem usually happens a few months after installation and get worse and more frequent with time.
Related to: Miilestone HF
The ZooMax HF and ConRad HF Technical Manuals incorrectly state that 1VDC on the filament feedback test point (BBU) refers to 1.3A filament current. This statement does not match the table below the statement with example values. The incorrect values are on page 58 and 60, respectively.
Related to: ConRad HF, ZooMax HF
One possible source of E14 errors is the electrical noise generated by the HV transformer primary cables. The noise may generate a false E14 when making a 200 mA or higher exposure...
Related to: ConRad HF, ZooMax HF
The ConRad / ZooMax generators do not have any potentiometers to adjust kV output. If necessary, the kV may be adjusted by software in Service Mode.
Related to: ZooMax HF